There are three models in the 67 series - Model 6700 covers the visible range of the spectrum and has a 4nm bandwidth; Models 6705 and 6715 cover the UV/visible range of the spectrum and have a 4nm and 1.5nm bandwidth respectively. All three instruments have modes for photometrics, spectrum scanning, multi-wavelength analysis, kinetics and quantitation, giving direct concentration results against single or multi-point calibrations. With extensive post measurement tools to ensure results are presented exactly to each users requirements the 67 series spectrophotometers also offer significant advances in data portability.
Jenway 6715
67 Series Spectrophotometers
- Colour touch screen display
- High quality, sealed split beam optics for optimum photometric performance
- Secure multi-user operation
- Save to SD memory cards
- Enhanced productivity
- Extensive range of accessories for flexibility and improve productivity
Together, the touch screen and colour display give the fastest, most flexible instrument interface possible. Where fine cursor control is required, using the QWheel makes the task easy and precise.
Jenway have developed Secure Multi User Operation for when instruments are shared by a number of users, so access to the instrument functions is controlled through a secure log-in procedure. For maximum security PIN codes can be allocated to individuals, groups or departments and to further increase flexibility each method can be given three levels of security:
Jenway have developed Secure Multi User Operation for when instruments are shared by a number of users, so access to the instrument functions is controlled through a secure log-in procedure. For maximum security PIN codes can be allocated to individuals, groups or departments and to further increase flexibility each method can be given three levels of security:
- Public access is free to all users.
- Read only methods can be accessed by all users, but only modified by the originator.
- Personal method can only be accessed by the logged in user
Technical Specification
Parameter | 6700, 6705 and 6715 |
Wavelength range | 320 to 1100nm (6700), 190 to 1100nm (6705, 6715) |
Wavelength resolution | 0.1nm |
Wavelength accuracy | ± 1.0nm |
Wavelength repeatability | ± 0.2nm |
Spectral bandwidth | 4nm (6700, 6705), 1.5nm (6715) |
Transmittance | 0 to 199.9%T |
Absorbance | - 0.300 to 3.000A |
Accuracy | ± 0.005A at 1.0A |
Resolution | 0.1%T, 0.001A |
Stability | < 0.001A/h after warm-up |
Quantitation range | - 99999 to +99999 |
Quantitation resolution | selec table 1, 0.1, 0.01 or 0.001 |
Calibration | Up to 20 standards with 5 replicates of each |
Units | mEq/l, ppm, mg/l, g/l, %, µg/ml, mg/ml, g/dl, mg/dl, µg/l, ng/l, µg/dl, M, mM, µM, U/l, mU/l, U/ml, blank |
Curve fit algorithms | Linear, quadratic and cubic functions |
Multi-wavelength | Up to 4 wavelengths |
Calculations | Ratio, difference, formulae with factors |
Kinetics time limits | 0 to 9999 seconds |
Calibration | Standard or factor |
Analysis | Curve details with mean rate of change plus formula of line of best fit |
Spectrum range | Any range between 320 (6700) or 190 (6705, 6715) and 1100nm |
Scan speed | 1500nm/min |
Scan interval | 0.1nm |
Analysis | Auto peaks and valleys, zoom, addition, subtraction, peak ratios, smoothing, area under curve, wavelength table, derivatives, analysis points, overlay |
Light source | Tungsten halogen lamp (6700), Xenon lamp (6705, 6715) |
Configuration | Secure multi-user and free access |
Number of users | 10 users + supervisor |
Methods/results memory | > 500 on internal flash memory or removable media |
Removable media | SD memory card |
Outputs | USB, Centronics, Analogue |
Size (w x d x h) | 490 x 390 x 220mm |
Weight | 7.5kg |
Görüntülenme (8040)
UV-VIS Spektrofotometreler
En Çok Bakılanlar
Son Baktıklarınız