The ADP440+ is single wavelength, high accuracy polarimeter suitable for use inmany applications and is especially suited for use in pharmaceutical laboratories where compliance with Pharmacopoeia¹ is required. New optics featuring a ‘no maintenance’ yellow LED and interference filter together with a photodiode detector allows reading of samples of up to 3.0 OD. The instrument has both angular and sugar scales (ISS) and can be programmed with user scales as well as industry standard methods for displaying concentration and specific rotation directly.
ADP 440+
ADP440+ - Operation
As the instrument is continually reading, all that is required is for the operator to place the sample tube in the measurement chamber and read the result. Four push buttons on the instrument keypad allow for MENU setup, ZERO calibration, METHOD selection (or display MODE if no methods are being used) and PRINT. A security system using passwords protects against deliberate or accidental misuse, which can be expanded to allow conformance with the technical requirements of FDA regulation 21 CFR Part 11. User Acess is provided by keypad entry or RFID user assigned ‘tags’.
The ADP440+ continually displays the measured reading, providing the operator assurance of measurement and sample stability. Data handling is in accordance with GLP providing a record of measurement date, time, batch and operator numbers.
On-screen menus in English, French, Spanish or German are selectable.
Where sample temperature is critical, optional polarimeter tubes fitted with a water jacket may be used in place of standard glass tubes. The instrument temperature sensor may be placed in the centre arm of the tube and if required, sugar or a user defined temperature compensation may be adopted.
ADP440+ – Sugar Purity Method
When used in the sugar industry, the ADP440+ may also form the central hub
of a purity system, providing the measurement of a Saccharimeter. In the
‘purity mode’ the instrument will calculate and display ‘sugar purity’ with Brix
values being input automatically from an RFM² refractometer or via the simple
keypad. For rapid measurements, funnel and flow through cells are available
and require the use of specially suited slotted lids.
ADP440+ - Calibration
The instrument can be checked and calibrated using a quartz control plate by
accessing the span facility in the setup menu. If a sample is too dark, the
instrument will display an error message.
1. Minimum requirement of EP/BP/JP/USP. 21CFR Part II
2. RFM700/340/800 only
Specifications | Angular | ISS (°Z) |
Range | -355 to + 355 selectable | -225 to +225 |
Resolution | 0.01/0.001 | 0.01/0.001 |
Reproducibility | 0.002 | 0.005 |
Accuracy | ±0.01 | ±0.03 |
Methods | Specific Rotation, Purity, % Concentration, % Invert Sugar, % inversion |
Light Source | LED/Interference filter (589nm) |
Beam Diameter | 4 mm |
Tube Length | 20 to 220 mm |
Reading Time | 4 – 30 seconds (selectable) |
Temperature | 5-40 °C |
Temperature Compensation | None, sugar, quartz, user defined |
Optical Density Range | 0.0 to 3.0 OD |
Interfaces | 2 x RS232, parallel printer port |
Power Supply | External 90-250V~, 50-60 Hz. Supplied with instrument |