The 72 series spectrophotometers are the first scanning spectrophotometers in the Jenway range to leverage diode array technology to produce exceptionally fast results. The range includes two models: model 7200 covers a wavelength range of 335 to 800nm with a spectral bandwidth of 7nm and model 7205 covers a wavelength range of 198 to 800nm with a spectral bandwidth of 5nm.
Jenway 7200
Both spectrophotometers are ideal for a variety of applications in education and routine testing in clinical, veterinary, pharmaceutical and QC laboratories.
Introducing the new 72 Series Scanning Spectrophotometers
• Scanning diode array technology
• Colour touchscreen navigation
• Small footprint and lightweight
• Fast scan speed
• English, French and German language options
• Multiple USB ports for data storage and printer connectivity
• Extensive range of accessories available
• 2 year warranty (including xenon lamp)
The 72 series spectrophotometers offer measurement modes for single wavelength with basic absorbance and % transmittance; concentration calculated using a standard or a factor; optical density; quantitation with the ability to create a calibration curve using up to 6 standards and 3 replicates; full spectrum scanning and kinetics where up to 3 wavelengths can be measured simultaneously.
The colour touchscreen user interface provides fast and easy set up and navigation of the instrument. The 4" display allows full spectrum scans, quantitation curves and kinetics runs to be viewed easily. The touchscreen capability enables users to zoom in and out and select spectral analysis points all by simply tapping the screen.
The large sample chamber will easily accommodate the optional long path length cuvette holder and the heated 10x10mm cuvette holder which is ideal for thermostated measurements at 37°C. The sample chamber lid can also be left open during measurements which is ideal for samples in tall test tubes.
There are two USB ports for data storage and printer connectivity. The easy access USB port on the front of the instrument can be used to easily store results and quantitation curves.
To take a tour round the intuitve user interface click here to use our simulator.
Technical Specification
Model | 7200 |
Wavelength Range | 335 to 800nm |
Wavelength Accuracy | ± 2nm |
Wavelength Repeatability | ± 2nm |
Spectral Bandwidth | 7nm |
Absorbance Range | - 0.300 to 2.500A |
Absorbance Accuracy | ± 0.01A at 1.0A and 546nm |
Absorbance Stability (A) | ± 0.005A/h at 0.04A and 546nm |
Noise | ± 0.002A at 0.1A and ± 0.02A at 2.0A and 546nm |
Stray Light at 340nm, %T | < 1%T according to ANSI/ASTM E387- 72 |
Concentration Range | ± 2500 |
Concentration Calibration | Blank with a single standard or factor |
Concentration Factor | ± 1000 |
Concentration Standard | ± 1000 |
Optical Density Factor | ± 1000 |
Quantitation Range | ± 2500 |
Quantitation Calibration | Blank with up to 6 standards |
Curve Fit Algorithms | Linear and linear through zero |
Kinetics Measurement Time | 15 to 9999 seconds |
Kinetics Number of wavelengths | 3 |
Kinetics Calibration | Blank with a factor |
Kinetics Display | Graphical, rate of change and concentration |
Kinetics Analysis | Concentration |
Spectrum Range | 335 to 800nm |
Spectrum Analysis | Absorbance or % transmittance and up to 50 spectral analysis points |
Beam Height | 15mm |
Light Source | Tungsten Halogen lamp |
Results Memory | Limited by attached USB memory stick |
Removable Media | USB |
Outputs | USB x 2 |
Supply voltage/frequency | 100 - 240VAC at 50 to 60Hz |
Power Supply | 12V DC, 3.8A |
Size (w x d x h) | 212 x 422 x 120mm |
Weight | 2.8kg |
Warranty | 2 years on the instrument, 1 year on the lamp |